Equipment

Element analyzer 2(C,H,N,S,O) | 원소분석기

보유장비 관련 정보
Laboratory/Field
Model Flash 2000
Maker Thermo
Technician Eun-jung Han
Contact 052-217-2640 / ejhan@unist.ac.kr
Status for Reservation 가능
Reservation Unit 1hr Maximum Time (per day) 12hr
Open(~ago) 5일전 Cancel (~ago) 2시간전
Equipment location 102동 B111호(Bldg.102, Room B111)
  • Description

    Element analyzer determinates the percentage eights of carbon, hydrogen, nitrogen, sulfur and oxygen of a sample. This information is important to help determine the structure of an unknown compound, as well as to help prove the structure and purity of a synthesized compound. It is based on the dynamic flash combustion of the sample followed by reduction, trapping, complete GC Separation and detection of the products by thermal conductivity detector.

  • Specifications

    • Measuring elements : C, H, N, S, O

    • Measuring range : 0.01 % ~ 100 %

    • Sample size : 0.01 ~ 500 mg

    • Accuracy : 0.1 ~ 0.3 % absolute

    • Detector : TCD

    • Furnace temperature : 1,100 °C

    • Analysis time

    - C, H, N, S : 10 min

    - oxygen : 5 min

  • Applications

    • Qualitative analysis of organic and inorganic compounds

    • Qualitative analysis of environmental samples

    • Qualitative analysis of geological samples Identification of empirical formula for organic compounds