Equipment

XPS | X선 광전자 분광기

보유장비 관련 정보
Laboratory/Field
Model K-alpha
Maker ThermoFisher
Technician Dong-ju Lim
Contact 052-217-4032 / djlim@unist.ac.kr
Status for Reservation 불가능
Reservation Unit 1hr Maximum Time (per day) 12hr
Open(~ago) 5일전 Cancel (~ago) 2시간전
Equipment location 울산광역시 울주군 언양읍 유니스트길 50, 102동 B117. 샘플보낼주소 : 울산시 울주군 언양읍 유니스트길 50 102동 203호
  • Description

    X-ray photoelectron spectroscopy (XPS) is a quantitative spectroscopic technique that measures the elemental composition, empirical formula, chemical state and electronic state of the elements that exist within a material. XPS spectra are obtained by irradiating a material with a beam of aluminum or magnesium X-rays while simultaneously measuring the kinetic energy (KE) and number of electrons that escape from the top 1 to 10 nm of the material being analyzed.

    의뢰 절차 및 참고사항(Analysis request procedure and sample preparation)
    1. 의뢰서 접수(장비의뢰, sample request 누른 후 양식 입력) 후 출력, 샘플과 함께 보낸다(시료보낼주소 참고)
    2. 샘플은 가로 세로 1cm x 1cm 정도로 자른다(높이는 2cm 이내에서 같은 높이로 잘라야 한꺼번에 진공으로 넣을수 있음)
    3. 파우더의 경우 실리콘기판(1cm x 1cm 정도) 혹은 알루미늄호일 위에 양면카본테이프를 붙이고 그 위에 스패츌러로 일정량의 파우더를 쏟고 꾹 눌러문지르세요. 그래서 양면테이프위에 덩어리진 파우더막이 형성되게 만들어주면 됩니다. 막의 너비는 3mm x 3mm 정도면 충분합니다. 다른 방법은 용매에 일정량을 섞어서 실리콘 기판(1cm x 1cm정도)위에 스핀코팅을 해서 두께 15nm 이상의 막을 형성시켜서 가져와도 됩니다.
    4. 샘플명은 가급적 일련의 숫자로 표기하기 바람(숫자가 싫다면 의뢰서 작성시 각각의 샘플이름을 타이핑해주길 바람)
    5. 외부 의뢰의 경우 샘플 도착하면 상태가 '분석의뢰중' 에서 '샘플접수'로 바뀝니다. 측정기간은 접수일 이후 10~15 일정도 소요됨
    6. 스케줄에 맞추어 샘플을 보내고 싶다면 의뢰서 접수후 따로 메일로 스케줄 요구해 줄 것을 바랍니다. 형평성 차원에서 10-15일 후에 일정을 잡게 됩니다.
    7. 외부 의뢰자의 경우 샘플 회수가 어려우니 가급적 샘플의 일부분만 보내줄것을 요청함(가로x 세로 : 3mm x 3mm 정도면 충분함)
    8. 1cm x 1 cm 의 샘플 크기일때 16개정도의 샘플을 한꺼번에 진공속으로 넣을수 있음
    9. 스케줄에 맞추어서 샘플 보낼때도 의뢰서 출력본을 함께 보내주세요.
    10. 앞에 언급한 측정 소요기간은 일반적인 경우에 그렇고 담당 행정업무나 다른 담당 장비에 시료가 몰릴경우 좀 더 늦춰질수 있습니다. 언급한 기간까지 만약 결과가 가지 않는다면 다시 한번 메일로 진행상황을 문의해주시면 고맙겠습니다.
    11. 시료의 전도성이 심하게 없다면(기준물질 PET 고분자정도의 전도성) charge compensation 이 잘 되지 않아서 결과가 적절하게 나오지 않을수 있습니다. 전도성물질과 비전도성물질이 복합되어있어도 결과가 만족스럽게 안 나올수 있습니다.
    12. 처음 의뢰하는 사람은 의뢰할 샘플이 아무리 많더라도 작은수의 샘플을 먼저 의뢰해서 결과를 받아본 후 만족스럽다면 그 후에 비슷한 종류의 샘플을 한꺼번에 많이 의뢰하는 것이 바람직합니다. 장비가 모든 샘플에 대해서 만족스런 결과를 못 낼수도 있습니다. 특히 전도성이 거의 없는 샘플들은 장비의 중화능이 샘플에 미치지 못할수 있으니 참고하기 바랍니다.
    12. 내부의뢰자의 결과는 항상 분석결과를 아래의 서버에 올려둡니다. 홈페이지에 완료로 표시되면 아래 사이트에서 결과를 다운 받을수 있습니다. 일반적으로는 자동으로 완료 메일이 의뢰자메일주소로 가게 되어있습니다만 어쩌다 한번씩 완료 메일을 못받았다는 사람이 있으니 참고바랍니다.
    13. 측정은 기본적으로 고진공하에서 이루어지므로 증기압을 발생하는 시료는 충분히 건조후 맡기기 바랍니다. 진공 잡을때 시간이 많이 소요되면 그에 맞추어서 장비사용료가 늘어납니다. 진공이 충분히 잡히지 않으면 측정결과 못 얻고도 사용료 청구 이루어질수 있습니다(상습적인 경우).
    14. 데이터 피팅 및 해석은 기본적으로 해주지 않습니다. 샘플 제조과정을 모르면 주관이 많이 개입하기 때문입니다.
    15. Tof-sims 측정항목은 Tof-sims 장비의뢰시에 선택하기 바랍니다. XPS 장비의뢰시는 체크 금지.
    16. UPS 스펙트럼을 얻는 것은 더이상 서비스하지 않습니다. 따라서 체크하지 말아주세요.
    17. 이 장비의 결과는 기본적으로 자동분석으로 하기 때문에 자체 알고리듬으로 언급한 원소 이외의 원소가 추가될 수 있습니다. 만약 그게 싫다면 수동으로 기재한 원소의 스펙트럼만 얻으라고 의뢰서에 기재를 해주세요.

    http://10.24.9.32:5000/
    위치: file station/Public/touser/각지도교수이니셜로 쓰인 폴더
    결과 파일 샘플이름 참조 : file station/Public/touser/SampleName

    샘플 보낼주소 : 울산시 울주군 언양읍 유니스트길 50 102동 203호


    Request Procedure and Important Notes
    1. Submission of Request
    After submitting a request form (click ‘Sample Request’) and completing all required information, print the request form and send it together with your samples (see the shipping address below).
    2. Sample Preparation (Solid Samples)
    Cut the sample into approximately 1 cm × 1 cm pieces. The height (thickness) should be less than 2 cm, and all samples should have a similar height so they can be loaded into the vacuum chamber simultaneously.
    3. Sample Preparation (Powder Samples)
    For powder samples:
    • Attach double-sided carbon tape to a silicon substrate (approximately 1 cm × 1 cm) or aluminum foil.
    • Place an appropriate amount of powder on the tape using a spatula.
    • Press and rub the powder firmly so that a compact powder film is formed on the tape.
    • A film size of approximately 3 mm × 3 mmis sufficient.
    Alternatively, the powder may be dispersed in a suitable solvent and spin-coatedonto a silicon substrate (approximately 1 cm × 1 cm) to form a film thickness greater than 15 nm.
    4. Sample Naming
    Please use a series of numbers as sample names whenever possible. If descriptive names are preferred, please enter the exact sample names in the request form.
    5. Status Updates for External Users
    For external requests, once the samples arrive, the status will change from "Request Submitted" to "Sample Received."
    The typical turnaround time is approximately 15 days after sample receipt.
    6. Scheduling Requests
    If you require analysis according to a specific schedule, please submit the request form first and then send a separate email requesting the desired schedule.
    To ensure fairness, scheduling is generally arranged at least 15 days after sample receipt.
    7. Sample Quantity for External Users
    Since returning samples may be difficult for external users, please send only a small portion of the material whenever possible.
    A sample size of approximately 3 mm × 3 mm is usually sufficient.
    8. Maximum Sample Capacity
    Approximately 16 samples of size 1 cm × 1 cmcan be loaded into the vacuum chamber simultaneously. The holder size is 5cm x 5cm. you cannot load the sample over 5cm x 5cm.
    9. Printed Request Form
    Even when samples are sent according to a prearranged schedule, please include a printed copy of the request form with the shipment.
    10. Possible Delays
    The turnaround time stated above applies under normal circumstances. Delays may occur due to administrative duties or a large number of samples awaiting analysis on this or other instruments.
    If you have not received results within the expected period, please contact us by email regarding the analysis status.
    11. Conductivity Requirements
    If a sample is highly insulating (e.g., having conductivity similar to or lower than PET polymer), charge compensation may not work effectively, and satisfactory results may not be obtained.
    Similarly, composite samples containing both conductive and non-conductive materials may produce unsatisfactory results.
    12. Recommendation for First-Time Users
    For first-time users, we strongly recommend submitting a small number of samples initially, regardless of the total number available.
    After reviewing the initial results, additional samples of a similar type may be submitted if the analysis is satisfactory.
    Please note that the instrument may not produce satisfactory results for every sample, particularly those with extremely low conductivity.
    13. Internal User Data Access
    Analysis results for internal users are uploaded to the following server:
    http://10.24.9.32:5000/
    Location:
    file station/Public/touser/[Advisor Initials Folder]
    Reference image for result file naming:
    file station/Public/touser/SampleName
    The folder contains a reference image showing how sample names appear in the analysis result files.


    When the request status is marked as Completed, the analysis results can be downloaded from the server.
    Completion notices are generally sent automatically to the requester's email address; however, occasional email delivery failures may occur. so you can check the server around 15 days after sample submission.
    14. Dry Samples Required
    All measurements are performed under high-vacuum conditions. Samples that generate vapor or outgassing products must be thoroughly dried before submission.
    Excessive pumping time will result in increased instrument usage charges.
    If sufficient vacuum cannot be achieved, analysis may fail while usage charges may still apply, particularly in cases of repeated submissions of unsuitable samples.
    15. Data Fitting and Interpretation
    Data fitting and detailed interpretation are not provided as part of the standard service because reliable interpretation often requires detailed knowledge of the sample preparation process.
    16. ToF-SIMS Requests
    Please select ToF-SIMS measurement items only when submitting a ToF-SIMS request.
    Do not select ToF-SIMS items when submitting an XPS request.
    17. UPS Measurements
    UPS (Ultraviolet Photoelectron Spectroscopy) measurements are no longer offered as a service.
    Please do not select UPS measurements on the request form.
    18. Automatic Element Identification
    XPS data processing utilizes automatic analysis algorithms. As a result, elements other than those specified in the request may be detected and included in the results.
    If only the manually specified elements should be analyzed, please clearly indicate this requirement in the request form.


    Sample Shipping Address
    XPS / ToF-SIMS FacilityRoom 203, Building 102UNIST (Ulsan National Institute of Science and Technology)50 UNIST-gil, Eonyang-eup, Ulju-gunUlsan 44919Republic of Korea

    Note:XPS and ToF-SIMS analyses are subject to sample-dependent limitations. Highly insulating, contaminated, wet, or otherwise vacuum-incompatible samples may not yield satisfactory results. We recommend submitting a small number of representative samples before committing to large-scale analysis requests.



    Sample Send Address: Unist-gil 502-dong 201-4, Unist-eup, Ulju-gun, Ulsan"

  • Specifications

    • Double-focusing hemispherical analyzer

    • Multi-element, high-transmission spectrometer input lens

    • 128-channel detector for high quality

    • Energy range : 200 eV to 3 keV

    • High flux even at low beam energy

    • 250 mm Rowland circle monochromator

  • Applications

    • Analysis of organic compound

    • Analysis of thin film

    • Depth profile