보유장비

GC x GC/TOFMS

보유장비 관련 정보
연구실/분야
모델명 Pegasus 4D GCxGC System
제조사 LECO
담당자
연락처 /
예약 가능여부 가능
예약안내 1시간 최대 예약시간 8시간
Open 없음 Cancel 없음
  • Description

    •GCxGC provides high peak capacity by trapping compounds eluting from the first chromatographic column and then systematically injecting them into the second column to obtain the second separation.
    • LECO's LN2 modulator delivers the ability to modulate these volatile analyses.
    •The Pegasus 4D, data acquisition rates of 100 to 500 spectra/second are commonly utilized, giving sufficient data density to obtain spectral deconvolution of the chromatographic peaks, using acquired data.

  • Specifications

    • Carrier Gas Flow : 0 to 10 mL/min Helium

    • LN2 Modulator Volatility Range : C4 to C40 n-alkane at amounts equal to or less than column sample capacity

    • LN2 Modulator Maximum Temperature : 400 °C

    • LN2 Modulator Maximum Heating Rate : 40 °C/min

    • Modulation Periods : 1 to 65 seconds

    • Ionization Source : Electron Impact

    • Mass Range : 5 ~ 1,000 amu

    • Spectral Generation Rate : 1 to 500 spectra/second

  • Applications

    • Petroleum

    • Geological

    • Environmental

    • Metabolomics

    • Food and Flavors

    • Pharmaceutical

    • Forensics

    • Identify of complex