GC/MS/MS | 가스크로마토그래피 질량분석기
연구실/분야 | |||
---|---|---|---|
모델명 | 450-GC & 320-MS | ||
제조사 | Bruker | ||
담당자 | 이경선 | ||
연락처 | 052-217-4036 / kslee@unist.ac.kr | ||
예약 가능여부 | 가능 | ||
예약단위 | 0.5hr | 1일최대예약시간 | 2hr |
예약Open(~일 전) | 5일전 | 예약취소불가(~일 전) | 1시간전 |
장비위치 | 102동 B111호(Bldg.102, Room B111) |
-
Description
GC/MS/MS makes an effective combination for chemical analysis. Mixture of volatile substances is separated by gas chromatography. Via a transfer line the fractions of each peak are transferred into the ion source where the ion ization takes place. After acceleration, the ions pass through a magnetic field which causes deviation of their linear flight curve according to their respective mass. After this magnetic field the detector counts the ions.
-
Specifications
• Ionization mode
- electron ionization
- positive/negative chemical ionization
• Mass range : 10 ~ 2,000 amu
• Mass axis stability : ± 0.1 unit over 24 hrs
• Resolution : 0.7 amu at 1,250 amu/sec,
0.6 amu at 500 amu/sec
• Mass type : tandem triple quadrupole
• Library : NIST library
-
Applications
• Confirmation of organic compounds structure
• Confirmation of molecular weight
• Identification of impurities and by-products